The Fluke 52 II dual-input digital thermometer delivers fast response with laboratory accuracy (0.05% + 0.3?C). You can use the 52 II to measure two contact temperature inputs simultaneously on motors, insulation, breakers, pipes, corroded connections, liquids, and wires with industrial standard J, K, T, or E-type thermocouple temperature sensors. The user friendly-front panel and large, backlit display make it easy to view results from both inputs in multiple ways. And the sleep mode feature preserves battery life to give you a typical thousand hours of operation.

  • Easy-to-access battery door that allows replacing the battery without breaking the calibration seal
  • Comes with splash- and dust-resistant case protected by an impact-absorbing holster and a three-year warranty
  • Powered by three, standard AA batteries -30-
  • Features an electronic offset function that compensates for thermocouple errors to maximize overall accuracy
  • Measures J, K, T, and E-type thermocouples
  • Handles temperatures up to 900?C (1600?F)
  • Presents results in ?C, ?F, or Kelvin (K)
  • Provides a user-friendly front panel that is easy to set up and operate
  • Offers sleep mode to increase battery life; 1000-hour battery life typical
  • Offers optional ToolPak accessory to hang the thermometer from any metal object (with the rare earth magnet) or secure around a pipe (with hook-and-loop straps) for hands-free operation
Specifications
Temperature accuracy Above -100?C J, K, T, E, and N-type: ?[0.05% + 0.3?C]1
R and S-type: ?[0.05% + 0.4?C]1
Below -100?C J, K, E, and N-types: ?[0.20% + 0.3?C]1
T-type: ?[0.50% + 0.3?C]
Temperature J-type -210?C to 1200?C
K-type -200?C to 1372?C
T-type -250?C to 400?C
E-type -150?C to 1000?C
N-type -200?C to 1300?C1
R and S-type 0?C to 1767?C1
Temperature scale ITS-90
Applicable standards NIST-175
Display resolution 0.1?C, 0.1 K < 1000
1?C, 1 K ? 1000
1. Only the Fluke Models 53 II B and 54 II Bthermometers are capable of measuring N, R, and S-type thermocouples.
Environmental Specifications
Operating temperature -10?C to 50?C
Storage temperature -40?C to 60?C
Humidity (without condensation) 0% to 90%; 0?C to 35?C
0% to 70%; 0?C to 50?C
Safety Specifications
Overvoltage category CSA C22.2 No. 1010.1 1992; EN 61010 Amendments 1,2
Agency approvals CE, CSA, T?V (pending)
Mechanical and General Specifications
Size (L x W x D) 173 x 86 x 38 mm
Weight 400 g
Batteries 3 AA batteries; typical 1000-hour life
Specification
Part No FLU52-2
Barcode 95969077644
Brand Fluke

Write a review

Captcha

Tags: Test And Measurement, Thermometers, Fluke, FLUKE Dual Probe Digital Thermometer FLU52-2